Operation
Table 4-4A. Chan # > Extensions > Corrections > Submenu (continued)
Menu Item Function
(Type) Selections (SCPI cmd)
Freq CF Pulse, Mod & Stat Display(or enter) the Frequency cal-factor
(Numeric) Continuous range:
3 dB This menu box displays the value of the current Frequency cal-factor.
This cal-factor is calculated by interpolation, if necessary, from a table
Meas
> Freq CH # menu box. This value can be overwritten using any of the
measurements until the Frequency is changed or the instrument is
initialized or re-started. At that time the interpolated value will be
restored.
SENSe[1|2]:CORRection:CALFactor < -3.00 to 3.00>
Temp Comp Pulse, Mod & Stat For temperature compensated sensors only, turns Temperature
Compensation On or Off
(Toggle) On, Off
This function defaults to On for sensors that have a valid temperature
compensation table. It may be turned Off for unusual testing situations.
For uncompensated sensors, this menu box does not appear and the
function is forced to Off.
SENSe[1/2]:CORRection:TEMPcomp {ON, OFF}
Table 4-4B. Chan # > Extensions > Define Pulse > Submenu
Menu Item Function
(Type) Selections (SCPI cmd)
Distal Pulse Sets the percentage of pulse amplitude defining the distal.
(Numeric) Continuous Range
The distal is normally defined as 90% of the pulse top level amplitude.
The range is from the 20% to 99.00%.
SENSe[1|2]:PULSe:DISTal <20.00 to 99.00> (percent)