Boonton 4500B RF Peak Power Analyzer
Maintenance
Calibrator External Pulse Verification
Verify the external pulse input for the calibrator after the sensor has been calibrated (Subsection
7.7). The external pulse input is a TTL-compatible input located on the rear panel of the Model
4500B. To test the external pulse input, proceed as follows:
1. Connect the Channel 1 sensor to the calibrator output.
2. Press the Init
3. Set the calibrator output to 5 dBm by selecting Calibrator > Set Level > 5 dBm.
4. Set the calibrator trigger to External by selecting Calibrator > Pulse > Source Ext.
5. Select positive trigger polarity by selecting Calibrator > Pulse > Polarity > +.
6. Set the calibrator to pulse mode by selecting Calibrator > Cal Mode Pulse.
7. Turn the calibrator on by selecting Calibrator > Cal Output > On.
8. Connect a BNC cable between the external pulse generator output and the EXT PULSE input
on the Model 4500B.
9. Set the external pulse generator output for 5 volts, 50 ohms, 100 ms pulse period, 50 ms pulse
width.
10. The Model 4500B should display a 5 dBm pulsed waveform with the period and pulse width
indicated in Step 9.