Boonton 4500B Peak Power Meter Manual de usuario Pagina 102

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 372
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 101
Boonton 4500B RF Peak Power Analyzer
Operation
4-30
Table 4-7. Chan Mem # > Menu
Menu Item Function
(Type) Selections (SCPI cmd)
Selection Pulse, Mod & Stat Identical to Chan # > menu. See Table 4-2.
See Table 4-2

Channel Pulse, Mod & Stat Turns the display of the current stored
(Toggle) Off, On waveform off or on.
Remove the stored waveform from the
waveform display window by pressing the
Chan Mem # > Channel menu key. Press
it again to restore the display.
CALCulate[6|7]:STATe {ON, OFF}
Vert Scale Pulse, Mod & Stat Initially identical to the Vertical Scale of
(Numeric) the waveform when saved.
The stored waveform is saved in full vertical
resolution in a floating point array. This
allows the vertical scale of the memory
channel waveform to be changed any time
after being recalled from memory or from
a file. The memory channel Units can also be
changed at any time using the Chan Mem #
> Exten > Units menu. Changes are
persistent, but do not modify the original file.
DISPlay:TRACe[6|7]:VSCALe <arg>
(current units)
Vert Center Pulse, Mod & Stat Initially identical to the Vertical Center
(Numeric) of the waveform when saved.
The stored waveform is saved in full vertical Figure 4-20.
resolution in a floating point array. This Chan Mem # >
allows the vertical center or offset of memory
channel waveforms to be changed any time
after being saved to memory or recalled from
a file. Changes are persistent, but do not
modify the original file.
DISPlay:TRACe[6|7]:VCENTer <arg> (current units)
Vista de pagina 101
1 2 ... 97 98 99 100 101 102 103 104 105 106 107 ... 371 372

Comentarios a estos manuales

Sin comentarios