Boonton 4500B RF Peak Power Analyzer
Operation
4-50
Table 4-9C. Trig > Menu (statistical)
Menu Item
(Type) Selections Function
Gating Stat Select free running or gated statistical mode
(Toggle) Free Run, Pulse Gated
Free Run use this mode to build a
distribution function that contains all of the
samples collected between the Start and
Stop measurement commands.
Pulse Gated use this mode to build a
distribution function that contains only the
samples within the time interval defined by
Gate 1 and Gate 2 that are collected between
the Start and Stop measurement commands.
This mode is used to measure the statistics
of a burst that can be synchronized with the
trigger and timebase system.
TRIGger:STATistical:GATIng {ON, OFF}
OFF = Free Run; ON = Pulse Gated
Trig Setup Gated Stat Setup the pulse gate trigger parameters
(Submenu) MENU
This menu is identical to Trig > menu above,
Table 4-9 and its submenu, Table 4-9A.
Trig > B Trigger. The paths will appear as
Trig > Trig A and Trig > Trig A > B Trig.
Stat* indicates a parameter that applies to
this setup.
Time Setup Gated Stat Setup the pulse gate timebase parameters
(Submenu) MENU
This menu is identical to the Time > menu
above, Table 4-8A. The path will appear as Fig. 4-27f. Trig > menu
the Trig > Time Settings > menu. Stat* Statistical Mode
indicates a parameter that applies to this
setup.
Gate 1 Gated Stat Mark the beginning of the statistical time gate
(Numeric) Continuous range:
With the burst of interest synchronized in the Pulse Preview window
position the Gate 1 vertical cursor at the start of the region to be
included in the distribution function.
TRIGger:STATistical:GATE[1] <arg> (sec)