Boonton 4500B RF Peak Power Analyzer
Operation
Table 4-12. Mark > Extensions > Submenu
Menu Item Function
(Type) Selections (SCPI cmd)
MK Group Pulse, Mod & Stat Selects the marker channel assignment mode.
(Toggle) Both, Each
are assigned to the same channel. Changing
either marker channel assignment changes
both
activates two choices in the Mark > Exten >
Delta Marker
can be assigned individually; that is, Mark >
Extensions > Mk 1 CH and Mark >
Extensions > Mk 2 CH can be assigned to
different channels. The marker channel
selections are independent and are saved
separately.
MARKer:GROUp {BOTH, EACH }
Marker Chan Pulse, Mod & Stat For MK Group both, assigns Marker 1 and
Mk 1 CH Active, CH1, CH2 Marker 2 to the same measurement channel.
Mk 2 CH Ext Trig 1, Ext Trig 2
(Mult. Choice) CH Math, For MK Group each, assigns Marker 1 and
Mem 1, Mem 2, Marker 2 to measurement channels.
When Active is selected, the Marker(s) is
assigned to and tracks the active channel.
in the Chan > Selection window.
The Time Mark # assignment for each of the
are independent of each other. When Fig. 4-29a.
Mark > Exten > Mk Group Both has been Mark > Exten menu
selected, both Marker 1 and Marker 2 are Both-Active
assigned to the same channel. Changing
either entry changes both.
When Mark > Mk Group Each has been selected, the marker
assignments are recalled from the last selections made in the Mark >
Mk Group Each mode. Each of the markers can be assigned to separate
measurement channels.