Boonton 4500B RF Peak Power Analyzer
Operation
4-65
Table 4-16. Meas > Param Meas > Submenu (pulse)
Menu Item Function
(Type) Selections (SCPI cmd)
Selection Pulse Select channel for status or parameter display
(Mult. Choice)
Param Mode Pulse Select type of display
(Toggle) Status, Meas
Select Status to display the following nine
parameters for the active channel in the field
above the grid box:
Freq, VidBW, Avging, TrA, TrB, Tr Dly,
Vscale, Vcent and Offset.
Select Meas to display nine selected
parameters in the display field above the
Grid box. The parameters are selected from
a list that includes the fifteen auto-measure
parameters as well as the status parameters.
The channel selection for each parameter is
determined by the active channel when the
selection is made. See Table 4-16A.
DISPlay:PARAmeter:MODE {MEAS,
STAT}
Param Column Pulse Select column in parameter display
(Mult. Choice) L, M, R
Param Top Pulse Select top item in column
(Mult. Choice) See Table 4-16A.
Param Middle Pulse Select middle item in column
(Mult. Choice) See Table 4-16A. Fig. 4-33.
Meas > Param Meas
menu
Param bottom Pulse Select bottom item in column
(Mult. Choice) See Table 4-16A.
DISPlay:PARAmeter:TOP:LEFT {see Table 5-1}
DISPlay:PARAmeter:MIDdle:LEFT {see Table 5-1}
DISPlay:PARAmeter:BOTtom:LEFT {see Table 5-1}
Middle and Right columns similar to above. See Chapter 5.