Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-54
5.5.11 OUTPut Subsystem
The OUTPut group of commands is used to control the internal 1GHz RF calibration source. The internal calibrator can be
configured as a programmable pulse source, with options for modulation rate and duty cycle. It can also be driven by an
external pulse modulation source.
OUTPut:LEVel:POWer
Description: Set or return the power level for the 1 GHz calibrator output signal.
Syntax: OUTPut:LEVel:POWer <asc>
Argument: <asc> = -50.0 to +20.0 dBm (0.1dB resolution)
Valid Modes: Any
OUTPut:LEVel:LIMit
Description: Set or return the limit power level in dBm for the 1 GHz calibrator output signal. The calibrator
output level cannot be set to exceed this limit.
Syntax: OUTPut:LEVel:LIMit <n>
Argument: <asc> = -40.0 to +20.0 (0.1dB resolution)
Valid Modes: Any
OUTPut:LEVel:STEP
Description: Set or return the power step size in dB used by the Knob and Arrow keys on the front panel to
control the output level of the 1 GHz calibrator.
Syntax: OUTPut:LEVel:STEP <n>
Argument: <asc> = 0.1 to 60.0 (0.1dB resolution)
Valid Modes: Any