Boonton 4500B RF Peak Power Analyzer
Operation
Table 4-43. Prgm > Trace Data > Menu
Menu Item Function
(Type) Selections (SCPI cmd)
Source See Chart Select a source for trace data
(Mult. Choice)
To select a source for trace data to save to a file, press the Prog >
TraceData > Source menu key to view the available sources:
Single Chan Dual Chan
CH1 CH1, CH2
Ext Trig 1 Ext Trig 1, Ext Trig 2
Mem1, Mem2 Mem1, Mem2
CH Math CH Math
+ Opt 10 + Opt 10
Cal Tbl 1 Cal Tbl 1, Cal Tbl 2
Histogram 1 Histogram 1, Histogram 2
TRACe:SOURce {CH1, CH2, TRIG1, TRIG2, CHM, MEM1,
MEM2, CAL1,CAL2, HIST1,HIST2}
Delimiter Comma, LF, Select an item delimiter character
(Mult. Choice) CR, Space
Press the Prog > Trace Data >Delimiter
menu key to select a character delimiter for
the trace file. The intended use of the data file
will determine your choice. The CR delimiter
is useful for the direct entry of the file into an
TRACe:DELImiter {COMMA, CR, LF, SPACE}
Basename Trace Opens the filename basename submenu
(Submenu)
Press the Prgm > Trace Data > Basename
menu key to open a menu that allows you to
change the basename of trace files. The
default basename is Trace. This can be
followed by a number that indicates the
stored sequence, e.g. Trace001.mmp,
Trace002.mmp, Trace003.mmp, etc.
An arbitrary basename may be used instead.
(None) DELETE Deletes the selected file.
(Action)
User confirmation is requested before a file is deleted.