Boonton 4500B RF Peak Power Analyzer
Operation
Table 4-51. Spcl >Trigger Out>Menu
Menu Item
(Type) Selections Function
Trig1 Source CH1, Trig1 Select the channel 1 rear panel BNC trigger out mode
(Mult. Choice) Sweep, 10 MHz
Clock
Trig1 Slope +, - Select rising or falling edge for Trig1 BNC out
(Toggle)
Trig2 Source CH2, Trig2 Select the channel 2 rear panel BNC trigger out mode
(Mult. Choice) Sweep, 10 MHz
Clock
Trig2 Slope +, - Select rising or falling edge for Trig2 BNC out
(Toggle)
OUTPut:TRIGout[1/2]:SOURce
{CH1. TRIG1, SWP, CLK for chan 1
CH2, TRIG2, SWP, CLK for chan 2}
OUTPut:TRIGout[1/2]:POLArity { POS, NEG}
Fig. 4-65.S pcl>Trig
Out menu
When TrigX Source is set to CH1 or CH2 is TTL signal which indicates the current state of that
its trigger comparator in real time. Any Low-to-High transition of this signal is a potentially triggerable edge, but does not
indicate that a sweep has actually been triggered.
When TrigX Source
sweep trigger. The signal will be low when the trigger is armed and waiting for a triggerable signal edge, and high from the
time the sweep is triggered until it has been acquired, processed, and the trigger re-arm criteria met for the next sweep,
including trigger holdoff and pre-trigger acquisition.
When TrigX Source is set to 10