Boonton 4500B RF Peak Power Analyzer
Remote Operation
DISPlay:PULSe:TIMEBASE
Description: Set or return the pulse mode timebase in seconds/division. The 4500B has fixed timebase settings
in a 1-2-5 sequence, and if the argument does not match one of these settings, it will be forced to
the next highest entry.
Syntax: DISPlay:PULSe:TIMEBASE <n>
Argument: <n> = 5e-9 to 10 sec/div in a 1-2-5 sequence,
30 sec/div,
1 to 10 min/div,
30 min/div and 60 min/div
Valid Modes: Pulse mode and Gated Statistical mode
Note: There are separate timebases for the Pulse/Gated Statistical mode and the Modulated mode.
The arguments selected are saved and restored independently by mode.
DISPlay:SCREensaver:BRIGhtness
Description: Set or return the display backlight brightness in percent when dimmed..
Syntax: DISPlay:SCREensaver:BRIGhtness <n>
Argument: <n> =10 to 50
Valid Modes: All
DISPlay:SCREensaver:STATe
Description: Set or return the state of the screensaver function. When ON, the screensaver dims the LCD
backlight to a specified intensity after a specified period of keyboard inactivity. GPIB activity
does not disturb display backlighting unless DISPlay:SCREensaver commands are sent.
Syntax: DISPlay:SCREensaver:STATe <asc>
Argument: <asc> = {ON, OFF}
Valid Modes: All
DISPlay:SCREensaver:TIME
Description: Set or return the time in minutes of no activity that dims the display backlight.
Syntax: DISPlay:SCREensaver:TIME <n>
Argument: <n> = 1 to 240, 999 disables the screensaver
Valid Modes: All