Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-63
READ:MARKer:MINimum?
Description: For the specified channel and marker, return the minimum power or voltage at the marker. The
units will be the same as the specified channel.
Syntax: READ[1|2|3|4|6|7]:MARKer[1|2]:MINimum?
Returns: CC, minimum power or voltage at marker
Where CC is the measurement condition code.
Valid Modes: Pulse and Modulated.
READ:MARKer:RATio?
Description: For the specified channel return the ratio of MK1 to MK2. The units will be dB for logarithmic
units or percent for linear units..
Syntax: READ[1|2|3|4|5|6|7]:Marker:RATio?
Returns: CC, MK1/MK2
Where CC is the measurement condition code.
Valid Modes: All
READ:MARKer:RDELTa?
Description: For the specified channel return the difference between MK2 and MK1. The units will be the same
as marker units.
Syntax: READ[1|2|3|4|5|6|7]:MARKer:RDELTa?
Returns: CC, (MK2-MK1)
Where CC is the measurement condition code.
Valid Modes: All
READ:MARKer:RRATio?
Description: For the specified channel return the ratio of MK2 to MK1. The units will be dB for logarithmic
units or percent for linear units..
Syntax: READ[1|2|3|4|5|6|7]:MARKer:RRATio?
Returns: CC, MK2/MK1
Where CC is the measurement condition code.
Valid Modes: All