Vista de pagina 255
Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-56
OUTPut:PULse:PERiod
Description: Set or return the pulse period in the Preset pulse control mode of the internal 1 GHz calibrator.
This parameter has no effect in the Variable pulse control mode External pulse source mode.
Syntax: OUTPut:PULse:PERiod <arg>
Argument: <arg> = {10e-3, 1e-3, 100e-6} seconds, or
{10ms, 1ms, 0.1ms}, or
{10000us, 1000us, 100us}
Valid Modes: 1 GHz Calibrator Preset Pulse control Mode
OUTPut:PULse:PERWID
Description: Set or return the pulse period <arg1> and pulse width <arg2> in seconds in the Variable pulse
control mode of the internal 1 GHz calibrator. This parameter has no effect in the Preset pulse
control mode or the External pulse source mode.
Syntax: OUTPut:PULse:PERWID <arg1> sp <arg2>
Argument: <arg1> = 28e-6 to 131.07e-3 (Pulse period)
<arg2> = 7e-6 to 65.535e-3 (Pulse width)
Valid Modes: Any
OUTPut:PULse:SOURce
Description: Set or return the pulse modulation source for the internal 1 GHz calibrator. If set to INT, the
            -panel EXT
PULSE input. Note this input requires TTL compatible signal levels.
Syntax: OUTPut:PULse:SOURce <asc>
Argument: <asc> = {INT, EXT}
Valid Modes: Any
OUTPut:SIGNal
Description: Set or return the on/off state of the internal 1 
Syntax: OUTPut:SIGNal <asc>
Argument: <asc> = {OFF, ON}
Valid Modes: Any.
Vista de pagina 255