Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-12
CALCulate:MATH:ARGA
Description: Set or return the signal source of the A argument for use in channel math. Channel math operates
pixel by pixel on two source channels to produce a third channel result
Syntax: CALCulate:MATH:ARGA <asc>
Argument: <asc> = {CH1 CH2, TRIG1, TRIG2, MEM1, MEM2}
Valid Modes: Any
CALCulate:MATH:ARGB
Description: Set or return the signal source of the B argument for use in channel math. Channel math operates
pixel by pixel on two source channels to produce a third channel result
Syntax: CALCulate:MATH:ARGB <asc>
Argument: <asc> = {CH1 CH2, TRIG1, TRIG2, MEM1, MEM2}
Valid Modes: Any
CALCulate:MATH:OPERator
Description: Set or return the pixel-by-pixel mathematical operation performed between sources A and B to
create the math channel result. Note that the apparent operation performed is different for
logarithmic and non-logarithmic units for equivalent results.
Syntax: CALCulate:MATH:OPERator <asc>
Argument: <asc> = {CH_SUM, CH_DIFF, CH_RAT}
Log units Non-log units
CH_SUM ------ add
CH_DIFF ------ subtract
CH_RAT subtract divide
Valid Modes: Any
CALCulate:MODE
Description: Set or return the system measurement mode. MODULATED is a continuous measurement mode,
PULSE is a triggered, oscilloscope-like mode, and STATISTICAL, which includes CDF, CCDF
and PDF as sub-modes, may be operated in a start-stop continuous mode, a decimated continuous
mode or a gated (triggered) mode.
Syntax: CALCulate:MODE <asc>
Argument: <asc> = {MODULATED, PULSE, STATISTICAL}
Note: The STATISTICAL sub-mode CDF, CCDF or PDF is restored but not reported.
Valid Modes: Any
Vista de pagina 211