Boonton 4500B Peak Power Meter Manual de usuario Pagina 208

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 372
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 207
Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-8
*OPC
Description: Clears the OPC (Operation Complete) status flag. This command is issued before the command to
be checked for completion. After this, the flag may be queried by *OPC? until a value of one is
returned, indicating the command has completed. Note that the query is not a true query - a value
of zero will never be returned.
Syntax: *OPC
Argument: None
Valid Modes: Any
*OPC?
Description:               
pending operations are complete. If pending operations are not yet complete, it does not return.
Syntax: *OPC?
Returns: Always returns 1 to indicate operations complete. Otherwise, does not return.
Valid Modes: Any
*OPT?
Description: Return the status of channel 1 and channel 2 followed by a list of installed options.
Syntax: *OPT?
Returns: f1 chan 1 installed?, f2 chan1 sensor present?, f3 chan 2 installed?, f4 chan 2 sensor
present?, opt1, opt2, etc. (option list may be empty).
Valid modes: Any
*RST
Description: Set the instrument to a known configuration. The current language selection (SCPI or 4500A
Emulation) is preserved. For SCPI, set measurements to STOP; for 4500A, set ,measurements to
RUN. Set the sensor temperature offset flag to FALSE, set the SCPI file over-write permission to
FALSE, turn the 1 GHz Calibrator output OFF and clear the error queues. System communication
parameters are not changed. Instrument measurement functions are set their default values (See
Table 3-3, Initialized Parameters).
Syntax: *RST
Argument: None
Valid Modes: Any
Vista de pagina 207
1 2 ... 203 204 205 206 207 208 209 210 211 212 213 ... 371 372

Comentarios a estos manuales

Sin comentarios