Boonton 4500B RF Peak Power Analyzer
Remote Operation
DISPlay:DECAdes
Description: Set or return the number of decades displayed on a Statistical mode (CDF, CCDF, PDF)
logarithmic horizontal scale.
Syntax: DISPlay:DECAdes <n>
Argument: <n> = 1 to 7
Valid Modes: Statistical modes only
DISPlay:ENVElope
Description: Set or return the status of the trace envelope display mode. Envelope mode plots the maximum
and minimum value for each measurement pixel.
Syntax: DISPlay:ENVElope <asc>
Argument: <asc> = {ON, OFF}
Valid Modes: Any
DISPlay:FULLscale
Description: Set or return the fullscale (right edge) value of the Statistical mode (CDF, CCDF, PDF)
logarithmic horizontal scale in percent.
Syntax: DISPlay:FULLscale <n>
Argument: <n> = {100, 10, 1, 0.1, 0.01}
Valid Modes: Statistical modes only
DISPlay:GRID
Description: Set or return the type of axes and graticule displayed in the traces box.
Syntax: DISPlay:GRID <asc>
Argument: <asc> = {BOX (box only), XHAIR (box with crosshair), GRID (box with grid), HASH (box with
hash-marks), XHAIRGRID (box with grid and crosshair)}
Valid Modes: All
Comentarios a estos manuales