Boonton 4500B Peak Power Meter Manual de usuario Pagina 246

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 372
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 245
Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-46
MARKer:MATH:EACH
Description: Set the marker group to EACH and configure the marker math function.
Syntax: MARKer:MATH:EACH <arg1>, <arg2>, <arg3>
Argument: <arg1> { RATIO, DELTA }
Specifies the result displayed in the center marker window
RATIO -the ratio of the left and right marker window readings in dBr or percent.
DELTA -valid only with <arg3> set to W (watts) or V (volts). The difference of the left
and right marker window readings expressed in channel units
<arg2> { MK1MK2, MK2MK1}
Specifies how the left and right marker windows are used to calculate the center marker
window.
MK1MK2 ratio is MK1/MK2; difference is MK1-MK2
MK2MK1 - ratio is MK2/MK1; difference is MK2-MK1
<arg3> Specifies the channel units.
{dBm, W, V, dBV, dBmV, dBuV}
Valid Modes: Modulated, Pulse and Statistical (RATIO with MK1MK2 only) modes
NOTE There is no query form for MARKer:MATH:BOTH or MARKer:MATH;EACH.
MARKer:POSItion:PERcent
Description: Set or return the percent probability (X-axis-position) of the selected vertical marker. Note that
the power value returned for each marker will depend on the setting of CALCulate:MODE.
When set to CDF, the highest power levels are towards the right side of the screen, with maximum
(highest peak) power occurring at 100%. When set to CCDF (also called 1-CDF), the highest
levels are towards the left, with peak power at 0%.
Syntax: MARKer[1|2]:POSItion:PERcent <n>
Argument: <n> = 0.000 to 100.000 %
Valid Modes: Statistical modes only.