Boonton 4500B RF Peak Power Analyzer
Operation
Table 4-17. Meas > Menu (Statistical)
Menu Item Function
(Type) Selections (SCPI cmd)
Freq Track Stat Selects the channel frequency entry mode.
(Toggle) On, Off
frequency is assigned to both channels.
Changing either CH1 or CH2 frequencies
changes both of them. Both windows are
updated.
frequency for each channel can be
individually assigned.
SENSe:CORRection:FRQTRK {ON, OFF}
Freq CH1 Stat Selects the input signal frequency on CH1
(Numeric) Continuous Range: for frequency response correction.
(Sensor Dependent)
See Table 4-15 above.
Freq CH2 Stat Selects the input signal frequency on CH2
(Numeric) Continuous Range: for frequency response correction.
(Sensor Dependent)
See Table 4- 15 above.
Sys Tracking Stat Controls the tracking of channel parameters.
(Toggle) Off, On
and video bandwidth track any changes made
on either channel.
Fig. 4-33a. Meas >
CALCulate:SYSTRK {ON, OFF} menu
Normalize Pwr Stat Select display and marker scaling of distributions
(Toggle) Off, On
isplayed as
normalized power versus probability. Each power point is divided by
the average power of all the points. Marker measurements also
change from power (dBm or watts) at a point to normalized power at
that point (dBr or percent).
SENSe:NORMalize {ON, OFF }