Boonton 4500B RF Peak Power Analyzer
Operation
4-116
NOTE The Channel 1 and Channel 2 voltage readings are derived by converting power to equivalent voltage.
Since power is non-polar, the derived voltage will always have positive polarity. Also, the average power
of a signal converted to voltage is not identical to the average voltage of that signal.
The External Trigger 1 and External Trigger 2 signals are always voltages and may be positive and
negative (bipolar). This 


the Pulse average voltage.
Table 4-58. Pulse Mode Automatic Measurements II
EXT TRIG 1&2
Name (Voltage Units)
Width Pulse width
Rise Pulse risetime
Fall Pulse falltime
Period Pulse Period
PRFreq Pulse repetition frequency
Duty C Pulse duty cycle
OffTime Pulse offtime
Peak Peak positive voltage
Pulse Pulse average voltage
Oversh Pulse overshoot (voltage ratio)
Avg Average voltage of pulse period
Top Pulse top level voltage
Bot Pulse bottom level voltage
Skew ----
EdgeDly Edge delay - time from left edge of
display to first mesial edge of either
slope on the displayed waveform