Boonton 4540 Series RF Power Meter
Operation
4-36
Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Corrections > Duty Cycle
Description: Sets the duty cycle in percent for calculated CW pulse power measurements. Valid only for
thermal sensors and CW sensors in the square-law region and subject to the accuracy of the duty
cycle value. Setting the duty cycle to 100% is equivalent to a CW measurement. Note that this
method of measuring pulse power should be used only if a peak power measurement cannot be
used. This item is not pictured aboveit only appears in Modulated mode. The Duty Cycle
Menu item is disabled for Peak Power Sensors.
Menu Type: Numeric Value
Range: 0.01% to 100.00%
SCPI Command: SENSe[1|2]:CORRection:DCYCle <numeric_value>
Fig. 4-29 Corrections
Menu (Statistical)
Voltage Probe
Fig. 4-30 Corrections
Menu (Statistical) Peak
Power Sensor - No TC
Fig. 4-31 Corrections
Menu (Statistical)
CW Power Sensor