Boonton 4540 Series RF Power Meter
Remote Operation
5-18
5.6.4 DISPlay Subsystem
The DISPlay group of commands is used to control the selection and presentation of textual, graphical and TRACe
measurements.
DISPlay:BACKlight:BRIGhtness
Description: Set or return the backlight brightness for the LCD display in percent. Zero percent turns the
backlight off.
Syntax: DISPlay:BACKlight:BRIGhtness <numeric_value>
Argument: <numeric_value> = 0 to 100
Valid Modes: All
DISPlay:CLEar
Description: Clear display traces and all data buffers for CH1 and CH2. Clears averaging filters to empty.
Does NOT clear errors.
Syntax: DISPlay:CLEar
Argument: None
Valid Modes: All
DISPlay:ENVELOPE
Description: Enable or disable the Envelope mode.
Syntax: DISPlay:ENVELOPE <Boolean>
Argument: <Boolean> = { 0, 1, ON, OFF }
Valid Modes: Pulse and Modulated
DISPlay:MODE
Description: Set or return the Text or Graphics page display setting. The Graphics page displays traces and
Marker measurements. The Text page displays auto-measure values in text form.
Syntax: DISPlay:MODE <character data>
Argument: <character data> = {GRAPH, TEXT}
Valid Modes: All