Boonton 4540 Series RF Power Meter
Remote Operation
5-89
Table 5-1. SCPI COMMAND SUMMARY (continued)
SENSe[1|2]:CORRection:CALFactor Set correction factor in dB. <numeric_value> = -3.00 to 3.00
SENSe[1|2]:CORRection:DCYCle Set/return duty cycle correction factor in percent.
<numeric_value> = 0.01 to 100.0%
SENSe[1|2]:CORRection:FREQuency Set channel frequency. <numeric_value> = 0.001e9 to
110.00e9 Hz
SENSe[1|2]:CORRection:OFFSet Set/return sensor offset value in dB. <numeric_value> = -300 to
300
SENSe[1|2]:CORRection:TEMPcomp Set/return temperature compensation. <Boolean> = { 0, 1, OFF,
ON }
SENSe[1|2]:FILTer:STATe Set/return filter state. <character data> = OFF, AUTO, ON
SENSe[1|2]:FILTer:TIMe Set or return the current length of the integration filter on the
selected channel. <numeric_value> = 0.002 to 16.000 seconds
SENSe[1|2]:IMPedance Set/return impedance reference for voltage to power in ohms.
<numeric_value> = 10.00 to 2500.00
SENSe[1|2]:PULSe:DISTal Set/return distal parameter of the risetime. <numeric_value> =
50.00 to 100.00
SENSe[1|2]:PULSe:ENDGT Set/return pulse gate end position in percent of pulse time duration.
<numeric_value> = 60.00 to 100.00
SENSe[1|2]:PULSe:MESIal Set/return mesial parameter of the risetime. <numeric_value> =
10.00 to 90.00
SENSe[1|2]:PULSe:PROXimal Set/return proximal parameter of the risetime. <numeric_value> =
0.00 to 50.00 percent
SENSe[1|2]:PULSe:STARTGT Set/return pulse gate start position in percent of pulse timeduration.
<numeric_value> = 0.00 to 40.00
SENSe[1|2]:PULSe:UNIT Set/return pulse definitions units. <numeric_value> = WATTS,
VOLTS
SENSe[1|2]:SENSOR:TYPE? Returns the sensor type for the selected channel. <character data>
= NONE, CW, VOLT, PEAK
SENSe:MBUF:SIZe Set/return size of measurement buffer for both channels.
<numeric_value> = -1 (FIFO), 0 (disabled), 1 to 1048576 points
SENSe:MBUF:RATe Set/return modulated mode meas buffer rate for both channels.
<numeric_value> = 1.0 to 1000.0 points per second
SENSe:MBUF:POSition? Returns the current meas buffer write index for both channels.