Boonton 4540 Series RF Power Meter
Remote Operation
5-21
DISPlay:[TEXt:]LIN:RESolution
Description: Set or return the display resolution for linear power and voltage readings. The number of
significant digits displayed is equal to the argument. This command also sets the resolution of
measurements returned in remote mode.
Syntax: DISPlay:[TEXt:]LIN:RESolution <numeric_value>
Argument: <numeric_value> = 3 to 5
Valid Modes: All
DISPlay:[TEXt:]LOG:RESolution
Description: Set or return the display resolution for logarithmic power and voltage readings. The number of
decimal places displayed is equal to the argument. This command also sets the resolution of
measurements returned in remote mode.
Syntax: DISPlay:[TEXt:]LOG:RESolution <numeric_value>
Argument: <numeric_value> = 0 to 3
Valid Modes: All
DISPlay:TRACe:HOFFSet
Description: Set or return the statistical mode horizontal scale offset in dB. The offset value will appear at the
leftmost edge of the scale with units dBr (decibels relative).
Syntax: DISPlay:TRACe[1|2]:HOFFSet <numeric_value>
Argument: <numeric_value> = -50.00 to +50.00 dB
Valid Modes: Statistical
DISPlay:TRACe:HSCALe
Description: Set or return the statistical mode horizontal scale in dB/Div.
Syntax: DISPlay:TRACe[1|2]:HSCALe <numeric_value>
Argument: <numeric_value> = 0.1 dB/Div to 5.0 dB/Div in a 1-2-5 sequence
Valid Modes: Statistical