Boonton 4540 Series RF Power Meter
Operation
4-22
Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Alarms > Low Limit
Description: Set the lower limit power level for the selected channel. When the measured average
power is below the lower limit, a down arrow will appear on the display above the
units on the main text screen in modulated mode, and flag bits are set in the remote mode
alarm register.
Menu Type: Numeric value.
Range: -300.000 to 300.000 dBm
SCPI Command: CALCulate[1|2]:LIMit:LOWer[:POWer] <numeric_value>