Boonton 4540 Series RF Power Meter
Remote Operation
Table 5-1. SCPI COMMAND SUMMARY (continued)
CALCulate[1|2]:PKHLD Set/return state of the peak hold function. <Boolean> = OFF,
AVG, INST
CALCulate[1|2]:REFerence:COLLect Set the current reading to be the ratiometric measurement reference
level.
CALCilate[1|2]:REFerence:DATA Set or return the ratiometric measurement reference level in dBm.
<numeric_value> = -200.00 to +200.00 dBm
CALCulate[1|2]:REFerence:STATe Set or return the ratiometric measurement mode state.
<Boolean> = 0, 1, OFF, ON
CALCulate[1|2]:STATe Enable currently selected channel allowing measurements to be
made. <Boolean> = 0, 1, OFF, ON
CALCulate[1|2]:UNIT s Change channel units. <numeric_value> = DBM, Watts, Volts,
DBV, DBMV, DBUV
CALibration[1|2]:[INTernal:]AUTO Start auto calibration with internal calibrator.
CALibration[1|2]:[INTernal:]FIXedcal Start fixed calibration with internal calibrator.
CALibration[1|2]:[INTernal:]ZERO Start zero calibration with internal calibrator.
CALibration[1|2]:EXTernal:AUTO Start auto calibration with external calibrator.
CALibration[1|2]:EXTernal:FIXedcal Start fixed calibration with external calibrator.
CALibration[1|2]:EXTernal:ZERO Start zero calibration with external calibrator.
DISPlay:BACKlight:BRIGhtness Set/return display brightness. <numeric_value> = 10 through 100
DISPlay:CLEar Clear measurement data and display
DISPlay:ENVELOPE Enable the Envelope display mode. <Boolean> = 0, 1, ON, OFF
DISPlay:MODE Place display into graphics or text mode. <character data> =
GRAPH, TEXT
DISPlay:MODUlated:TIMEBASE Set/return modulated timebase.
DISPlay: MODUlated:TSPAN Set/return horizontal time span.Ex: <numeric_value> = 0.05 (50
ms)
DISPlay:PULSe:TIMEBASE Set/return horizontal timebase. Ex: <numeric_value> = 0.05 sets
timebase to 50 ms
DISPlay:PULSe:TSPAN Set/return horizontal time span. Ex:<numeric_value> = 0.05 (50
ms)
DISPlay:SCREensaver:BRIGhtness Set/return display brightness in screensaver mode.
<numeric_value> = 10 to 50