Boonton 4540 Series RF Power Meter
Operation
4-28
Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Define Pulse > Pulse Units
Description: Set the units for entering the pulse distal, mesial, and proximal levels. If Volts are
selected, the pulse transition levels will be defined as the specified percentage in voltage.
If set to WATTS, the levels are defined in percent power. Many pulse measurements call
for 10% to 90% voltage (which equates to 1% to 81% power) for rise time and falltime
measurements, and measure pulse widths from the half-power (3dB, 50% power, or
71% voltage) points.
Menu Type: Toggle.
Choices: Watts, Volts
SCPI Command: SENSe[1|2]:PULSe:UNIT <character data>
Main > Channel > 1 | 2 >Extensions Pg 1 > Define Pulse > Pulse Gate
Description: Selects the Pulse Gate menu.
Menu Type: Menu Selection
SCPI Command: None