Boonton 4540 Series RF Power Meter
Remote Operation
OUTPut:MIO:TRIGout:SOURce
Description: Set/return Multi IO source for trigger out mode. Below is a description of each argument:
Argument Description
INT1, INT2: Multi IO output is active when selected internal trigger is active.
EXT: Multi IO output is active when external trigger is active.
SWEEP: Pulses when actual trigger event occurs.
Syntax: OUTPut:MIO:TRIGout:SOURce <character data>
Argument: <character data> = { INT1, INT2, EXT, SWEEP}
Valid Modes: Any
OUTPut:MIO:VOLTage
Description: Set/return Multi IO user set voltage out.
Syntax: OUTPut:MIO:VOLTage:SOURce <numeric_value>
Argument: < numeric_value > = -10.0 to + 10.0 V
Valid Modes: Any