5-1. SCPI COMMAND SUMMARY (continued)
FETCh[1|2]:INTERval:PKAVG? Return peak to average power between MK1 and MK2.
FETCh:KEY? Return code of last key depressed
FETCh[1|2]:MARKer[1|2]:AVERage? Return current reading at the specified marker.
FETCh[1|2]:MARKer:DELTa? Return difference between MK2 and MK1.
FETCh[1|2]:MARKer[1|2]:MAXimum? Return maximum reading at the specified marker
FETCh[1|2]:MARKer[1|2]:MINimum? Return minimum reading at the specified marker
FETCh[1|2]:MARKer:CURsor:PERcent? Return percent at cursor.
FETCh[1|2]:MARKer:CURsor:POWer? Return relative power at cursor.
FETCh[1|2]:MARKer:RATio? Return ratio between MK2 and MK1 as a percentage.
FETCh[1|2]:MARKer:RDELTa? Return difference between MK1 and MK2.
FETCh[1|2]:MARKer:RRATio? Return ratio between MK1 and MK2 as a percentage.
FETCh[1|2]:TEMPerature:AVERage? Return average sensor internal temperature in degrees C
FETCh[1|2]:TEMPerature:CURRent? Return instantaneous sensor internal temperature in degrees C
FETCh:TEMPerature:INTernal? Return current internal instrument temperature in degrees C
INITiate:CONTinuous Set/return state of mode which triggers meas cycles continuously.
<Boolean> = 0, 1, OFF, ON
INITiate[:IMMediate[:ALL]] Set mode which starts a measurement cycle when trigger event
occurs
MARKer[1|2]:POSItion:PIXel Set/return marker position in pixels for selected window.
<numeric_value> = 0 through 500
MARKer[1|2]:POSItion:TIMe Set/return marker time relative to the trigger.
MARKer:POSItion:PERcent Set or return the percent probability (y-axis-position) of the CCDF
cursors.<numeric_value> = 0.000 to 100.0 %
MARKer:POSItion:POWer Set or return the cumulative relative power (x-axis-position) of the
CCDF cursors. <numeric_value> = 0.000 to 200.000
MEASure[1|2]:POWer? Return Modulated Mode Pavg in dBm
MEASure[1|2]:VOLTage? Return Modulated Mode Pavg in equivalent volts.