4-53
Boonton Electronics Chapter 4
4530 Series RF Power Meter Remote Operation
*ESE
Description: Set or return the Standard Event Status Enable Register. The mask value in this register is
used to enable particular bits for generating a service request (SRQ) over the GPIB when
certain conditions exist in the Event Status Register. When a mask bit is set, and the
corresponding ESR bit goes true, an SRQ will be generated, provided the Event Status
Summary bit (ESB, bit 5) is enabled in the SRE register. No SRQ can be generated for that
condition if the mask bit is clear. See the *ESR command for bit assignments.
Syntax: *ESE <n>
Argument: <n> = 0 to 255
Valid Modes: Any
*ESR
Description: Return the current value of the Standard Event Status Register, then clear the register. This
register has bits assigned to a number of possible events or conditions of the instrument.
The register value may be read using this command, or may be used to generate a service
request (SRQ) over the GPIB when certain conditions exist. Individual bits may be enabled
or disabled for SRQ generation using the ESE mask (see *ESE command). The following
table shows the bit assignments in the Standard Event Status Register:
Bit Value Definition
0 1 OPeration Complete flag 1 = all current operations have completed execution.
1 2 Not used always returns 0
2 4 Not used always returns 0
3 8 Device Dependent Error 1 = the instrument encountered a device dependent error.
4 16 Execution Error 1 = the instrument encountered an execution error.
5 32 Command Error 1 = a remote interface command error exists.
6 64 Not used always returns 0
7 128 PON 1 = the instrument power has been turned on.
Syntax: *ESR?
Returns: Current Value of Event Status Register (0 to 255)
Valid Modes: Any
*IDN
Description: Return the instrument identification string. This string contains the manufacturer, model
number, serial number and firmware version number.
Syntax: *IDN?
Returns: < Mfgr, Model#, Serial#, Version >