8.1 File
8.1.1 Start / Stop History Save
This feature allows users to log measurements over a period of time in Analyzer
mode. The log files are stored in TSV (Tab Separated Value) ASCII text format.
Values listed are:
Date, Time, Carrier1 (MHz), Carrier2 (MHz), Output Level1 (dBm), Output Level2
(dBm), Offset (dB), ALC On/Off, Screen Mode, IM Number (Order), IM
Frequency (MHz), IM Measure(dBc), Direction, Measure Mode/Pass IM
Level/decision
Note: PIM 31 does not support Offset, Measure Mode/Pass and IM
Level/decision. This are reserved for other test systems. The log data will show a
hyphen “-“in these columns. Direction is always reverse for PIM31 test systems.
Each log cycle requires manual start, entering a file name, and manual stop. File
name directory and drive can be chosen at liberty.