13.3 Performance Verification
Performance Verification measures output signals and frequency accuracy /
selectivity of the PIM31.
13.3.1 Tx Signal Power Level
The following components are required for the performance verification.
50W Low-PIM Termination Load (Boonton 95960501A)
Low PIM Cable (Boonton: 95960401A)
High Power coupler 30dB coupling (with known linearity)
RF Power Meter (Boonton 4540 w/ Sensor 51071A)
Install the test set-up as shown:
Vista de pagina 85