Boonton 9240 Series RF Voltmeter
STATus:OPERation:NTRansition
Description: Set or return the value of the negative transition filter bitmask for the Operation status group.
Setting a bit in the negative transition filter causes a 1 to 0 (negative) transition in the
corresponding bit of the Operation Condition register to cause a 1 to be written in the associated
bit of the Operation Event register.
Syntax: STATus:OPERation:NTRansition <numeric value>
Argument: <numeric value> = 0 to 65535
Returns: 16-bit register value (0 to 65535)
STATus:OPERation:PTRansition
Description: Set or return the value of the positive transition filter bitmask for the Operation status group.
Setting a bit in the positive transition filter causes a 0 to 1 (positive) transition in the
corresponding bit of the Operation Condition register to cause a 1 to be written in the associated
bit of the Operation Event register.
Syntax: STATus:OPERation:PTRansition <numeric value>
Argument: <numeric value> = 0 to 65535
Returns: 16-bit register value (0 to 65535)
STATus:PRESet
Description: Sets SCPI enable registers and transition filters to the default state. The Operational Enable and
Questionable Enable mask registers are both cleared so an SRQ will not be generated for these
conditions. All bits for the device and questionable calibration registers are enabled. All positive
transition filters are enabled and all negative transition filters are cleared.
Syntax: STATus:PRESet