Boonton 9240 Series RF Voltmeter
IEEE Menu. The IEEE submenu is used to configure the Model 9240 for communications over the GPIB. An example of the
menu is shown in Figure 3-9 and the description of the commands, parameters and factory defaults is given in Table 3-4.
R T N
A D D R E S S > 1 3
E M U L A T I O N S C P I
E O S L S T N C R L F
E O S T A L K E R C R L F
E O I O N
S R Q M A S K 0
Figure 3-9. IEEE Menu Display
Table 3-4 IEEE MENU Functions
Function Description Parameters Defaults
RTN Returns the instrument to the n/a
previous menu.
ADDRESS GPIB address assigned to the 0 to 30 n/a
instrument.
EMULATION GPIB emulation mode. SCPI, 437B, 438A, 4230 SCPI
EOS LSTN End of string indicator for LF, CR, CRLF, NONE LF
received messages. Where:
LF = Line Feed
CR = Carriage Return
CRLF = Carriage Return
and Line Feed
EOS TALKER End of string character sent LF, CR, CRLF, NONE LF
with transmitted messages.
EOI Enables/disables the end or ON, OFF OFF
identify hardware control line.
SRQ MASK Service request interrupt mask. 0 to 255 0
See Section 5.5.1 *STB? for bit Where:
descriptions. 255 enables all interrupts
Getting Started
3-12