4540 Power Meter
Quick Start Guide
PN: 98406000A © Copyright 2008 Boonton Page 37
4540 RECOMMENDED SENSORS - PEAK
Sensor Response Maximum SWR
Model
Impedance
RF
Connector
Frequency
Range
(Low
Bandwidth)
Dynamic
Range
Peak Pwr Rng
CW Pwr Rng
Int Trigger Rng
Overload
Rating
Pulse /
Continuous
Fast
Rise Time
(Bandwidth)
Slow
Rise Time
(Bandwidth)
Frequency
SWR
57318
50 ohm
N (M)
0.5 - 18 GHz
(0.05 - 18
GHz)
-24 to +20 dBm
-34 to +20 dBm
-10 to +20 dBm
1 W for 1µs
200 mW
<15 ns(2)
(35 MHz)
<10 µs
(350 kHz)
0.05 -
2 GHz
2 - 6 GHz
6 - 16 GHz
16 - 18 GHz
1.15
1.20
1.28
1.34
57340
50 ohm
K (M)
0.5 - 40 GHz
(0.05 - 40
GHz)
-24 to +20 dBm
-34 to +20 dBm
-10 to +20 dBm
1 W for 1µs
200 mW
<15 ns(2)
(35 MHz)
<10 µs
(350 kHz)
0.05 -
4 GHz
4 - 38 GHz
38 - 40 GHz
1.25
1.65
2.00
57518
50 ohm
N (M)
0.1 - 18 GHz
(0.05 - 18
GHz)
-40 to +20 dBm
-50 to +20 dBm
-27 to +20 dBm
1 W for 1µs
200 mW
<100 ns
(6 MHz)
<10 µs
(350 kHz)
0.05 -
2 GHz
2 - 6 GHz
6 - 16 GHz
16 - 18 GHz
1.15
1.20
1.28
1.34
57540
50 ohm
K (M)
0.1 - 40 GHz
(0.05 - 40
GHz)
-40 to +20 dBm
-50 to +20 dBm
-27 to +20 dBm
1 W for 1µs
200 mW
<100 ns
(6 MHz)
<10 µs
(350 kHz)
0.05 -
4 GHz
4 - 38 GHz
38 - 40 GHz
1.15
1.65
2.00
Please note:
Additional sensors are possible with the use of an external 1GHz Calibration
source.
For further requirements please contact your next Boonton support engineer.