Boonton 4240 RF Power Meter Manual de usuario Pagina 83

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 128
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 82
Boonton 4240 Series RF Power Meter
5.5.10 READ Queries
he purp AD
ata. READ s the
is equivalent to ABORtin
ready. READ? generally
every time it is issued, RE
this case, use FETCh que
SENSe:FILTer) is filled.
r?
Description: Return current average amplitude reading in channel units.
READ[1|2]:CW:POWer?
T ose of the RE ? group of queries is to initiate a measurement cycle, acquire data, and return specific measurement
d ? perform initiation, data acquisition, postprocessing, and data output portions of the measurement. READ?
g any operation in progress, INITiating a new measurement, then FETChing the data when it is
does not return data unless acquisition is complete. Since READ? INITiates a new measurement
AD? queries should not be used for free running data acquisition (INITiate:CONTinuous ON) - in
ries instead. The measurement is generally considered complete when the integration filter (see
READ:CW:POWe
Syntax:
Returns: CC, Average power (watts, dBm)
Where CC is the measurement condition code.
Remote Operation
5-23
Vista de pagina 82
1 2 ... 78 79 80 81 82 83 84 85 86 87 88 ... 127 128

Comentarios a estos manuales

Sin comentarios